Remote Commands
R&S
®
FSW
819User Manual 1173.9411.02 ─ 19
Return values:
<PulsePower> List of pulse powers.
The number of values depends on the number of pulses you
have been measuring.
The unit is dBm.
Usage: Query only
[SENSe:]MPOWer:[:SEQuence] <Frequency>, <RBW>, <MeasTime>,
<TriggerSource>, <TriggerLevel>, <TriggerOffset>, <Detector>, <#OfPulses>
This command configures and initiates the pulse power measurement.
The R&S FSW caches all measurement parameters that you can set with this com-
mand. If you use the command repeatedly, the R&S FSW only changes those settings
that you have actually changed before initiating the measurement. Thus, measurement
times are kept as low as possible.
If you synchronize the measurement with *OPC, the R&S FSW produces a service
request when all frequencies have been measured and the number of individual meas-
urements has been performed.
Note that using the command as a query initiates the measurement and returns the
results if all frequencies have been measured. For more information on querying the
results see [SENSe:]LIST:POWer:RESult?.
Parameters:
<Frequency> Defines the pulse frequency.
Range: 0 to Fmax
Default unit: Hz
<RBW> Defines the resolution bandwidth.
<MeasTime> Defines the measurement time.
Range: 1 µs to 30 s
<TriggerSource> EXTernal | EXT2 | EXT3 | VIDeo
Selects a trigger source.
For more information see Configuring Triggered and Gated
Measurements.
<TriggerLevel> Defines a trigger level.
The trigger level is available for the video trigger. In that case,
the level is a percentage of the diagram height.
In case of an external trigger, the R&S FSW uses a fix TTL level.
Range: 0 to 100
Default unit: PCT
<TriggerOffset> Defines the trigger delay.
Range: 0 s to 30 s
*RST: 0 s
Default unit: s
Configuring and Performing Measurements