Remote Commands
R&S
®
FSW
842User Manual 1173.9411.02 ─ 19
Note that the number of sweep points is limited to 10001 when measuring spurious
emissions.
For EMI measurements, 200001 sweep points are available.
Parameters:
<SweepPoints> Range: 101 to 100001
*RST: 1001
Example:
SWE:POIN 251
Usage: SCPI confirmed
Manual operation: See "Sweep Points" on page 413
[SENSe:]SWEep:TIME <Time>
This command defines the sweep (or: data capture) time.
In the Spectrum application, the command decouples the sweep time from the span
and resolution and video bandwidths.
Parameters:
<Time> refer to data sheet
*RST: depends on current settings (determined automati-
cally)
Example:
SWE:TIME 10s
Usage: SCPI confirmed
Manual operation: See "Sweep Time" on page 155
See "Sweep Time" on page 411
See "Sweep Time" on page 427
[SENSe:]SWEep:TIME:AUTO <State>
This command couples and decouples the sweep time to the span and the resolution
and video bandwidths.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
SWE:TIME:AUTO ON
Activates automatic sweep time.
Usage: SCPI confirmed
Manual operation: See "Harmonic Sweep Time" on page 274
See "Sweep Time" on page 411
See "Default Coupling" on page 412
Setting Basic Measurement Parameters