Measurements
R&S
®
FSW
274User Manual 1173.9411.02 ─ 19
monic Distortion" dialog box, which is displayed as a tab in the "Analysis" dialog box or
when you select the "Harmonic Distortion Config" softkey from the "Harm Dist" menu.
The remote commands required to perform these tasks are described in
chap-
ter 11.5.10, "Measuring the Harmonic Distortion", on page 795.
No. of Harmonics........................................................................................................ 274
Harmonic Sweep Time................................................................................................274
Harmonic RBW Auto...................................................................................................274
Adjust Settings............................................................................................................ 275
No. of Harmonics
Defines the number of harmonics to be measured. The range is from 1 to 26. Default is
10.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:NHARmonics on page 797
Harmonic Sweep Time
Defines the sweep time for the zero span measurement on each harmonic frequency.
This setting is identical to the normal sweep time for zero span, see also "Sweep Time"
on page 411.
Remote command:
[SENSe:]SWEep:TIME:AUTO on page 842
Harmonic RBW Auto
Enables/disables the automatic adjustment of the resolution bandwidth for Normal
(3dB) (Gaussian) and 5-Pole filter types. The automatic adjustment is carried out
according to:
"RBW
n
= RBW
1
* n"
If RBW
n
is not available, the next higher value is used.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:HARMonics:BANDwidth:AUTO
on page 796
Harmonic Distortion Measurement