Remote Commands
R&S
®
FSW
861User Manual 1173.9411.02 ─ 19
MSRA operating mode
In MSRA operating mode, settings related to data acquisition (measurement time, hys-
teresis) can only be adjusted automatically in the MSRA Master, not in the MSRA
applications.
[SENSe:]ADJust:ALL......................................................................................................861
[SENSe:]ADJust:CONFigure:DURation............................................................................ 861
[SENSe:]ADJust:CONFigure:DURation:MODE..................................................................862
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer..............................................................862
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer...............................................................862
[SENSe:]ADJust:CONFigure:TRIG...................................................................................863
[SENSe:]ADJust:FREQuency..........................................................................................863
[SENSe:]ADJust:LEVel...................................................................................................863
[SENSe:]ADJust:ALL
This command initiates a measurement to determine and set the ideal settings for the
current task automatically (only once for the current measurement).
This includes:
Center frequency
Reference level
Example:
ADJ:ALL
Usage: Event
Manual operation: See "Adjusting all Determinable Settings Automatically (Auto
All)" on page 437
[SENSe:]ADJust:CONFigure:DURation <Duration>
In order to determine the ideal reference level, the R&S FSW performs a measurement
on the current input data. This command defines the length of the measurement if
[SENSe:]ADJust:CONFigure:DURation:MODE is set to MANual.
Parameters:
<Duration> Numeric value in seconds
Range: 0.001 to 16000.0
*RST: 0.001
Default unit: s
Example:
ADJ:CONF:DUR:MODE MAN
Selects manual definition of the measurement length.
ADJ:CONF:LEV:DUR 5ms
Length of the measurement is 5 ms.
Manual operation: See "Changing the Automatic Measurement Time (Meastime
Manual)" on page 438