Remote Commands
R&S
®
FSW
882User Manual 1173.9411.02 ─ 19
Suffix:
<p>
.
1 | 2 | 3
Selects the connector:
1 = Baseband Input I
2 = Baseband Input Q
3 = RF (currently not supported; use "1" with RF Input Connec-
tor setting "Baseband Input I")
Parameters:
<Mode> RSINgle
Run single: starts one data acquisition.
NOACtion
Nothing is started on pressing the micro button.
*RST: RSINgle
Manual operation: See "Microbutton Action" on page 321
[SENSe:]PROBe<p>:SETup:NAME?
Queries the name of the probe.
Suffix:
<p>
.
1 | 2 | 3
Selects the connector:
1 = Baseband Input I
2 = Baseband Input Q
3 = RF (currently not supported; use "1" with RF Input Connec-
tor setting "Baseband Input I")
Return values:
<Name> Name string
Usage: Query only
[SENSe:]PROBe<p>:SETup:STATe?
Queries if the probe at the specified connector is active (detected) or not active (not
detected). To switch the probe on, i.e. activate input from the connector, use
INP:SEL:AIQ (see INPut:SELect on page 866).
Suffix:
<p>
.
1 | 2 | 3
Selects the connector:
1 = Baseband Input I
2 = Baseband Input Q
3 = RF (currently not supported; use "1" with RF Input Connec-
tor setting "Baseband Input I")
Return values:
<State> DETected | NDETected
*RST: NDETected
Setting Basic Measurement Parameters