Remote Commands
R&S
®
FSW
938User Manual 1173.9411.02 ─ 19
In the spectrogram, the vertical marker position remains the same.
Usage: Event
Manual operation: See "Search Mode for Next Peak" on page 499
See "Search Mode for Next Peak in X Direction" on page 502
See "Search Next Peak" on page 506
CALCulate<n>:MARKer<m>:MAXimum[:PEAK]
This command moves a marker to the highest level.
In the spectrogram, the command moves a marker horizontally to the highest level in
the currently selected frame. The vertical marker position remains the same.
If the marker is not yet active, the command first activates the marker.
Usage:
Event
Manual operation: See "Marker Search Type" on page 503
See "Peak Search" on page 506
CALCulate<n>:MARKer<m>:MAXimum:RIGHt
This command moves a marker to the next lower peak.
The search includes only measurement values to the right of the current marker posi-
tion.
In the spectrogram, the vertical marker position remains the same.
Usage:
Event
Manual operation: See "Search Mode for Next Peak" on page 499
See "Search Mode for Next Peak in X Direction" on page 502
CALCulate<n>:MARKer<m>:MINimum:AUTO <State>
This command turns an automatic marker peak search for a trace minimum on and off.
The R&S FSW performs the peak search after each sweep.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:MARK:MIN:AUTO ON
Activates the automatic minimum value search function for
marker 1 at the end of each particular sweep.
Manual operation: See "Auto Max / Min Peak Search" on page 500
CALCulate<n>:MARKer<m>:MINimum:LEFT
This command moves a marker to the next minimum value.
Analyzing Measurements (Basics)