Remote Commands
R&S
®
FSW
956User Manual 1173.9411.02 ─ 19
11.8.3.9 Noise Measurement Marker
The following commands control the noise measurement marker function.
CALCulate<n>:MARKer<m>:FUNCtion:NOISe:RESult?.....................................................956
CALCulate<n>:MARKer<m>:FUNCtion:NOISe[:STATe].....................................................956
CALCulate<n>:MARKer<m>:FUNCtion:NOISe:RESult?
This command queries the result of the noise measurement.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps. See also INITiate:CONTinuous on page 694.
Return values:
<NoiseLevel> Current noise level. The unit is the one currently active.
Example:
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK2 ON
Switches on marker 2.
CALC:MARK:FUNC:NOIS ON
Switches on noise measurement.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK2:NOIS:RES?
Outputs the noise result of marker 2.
Usage: Query only
Manual operation: See "Noise Measurement State" on page 510
CALCulate<n>:MARKer<m>:FUNCtion:NOISe[:STATe] <State>
This command turns the noise measurement at the marker position on and off.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CALC:MARK:FUNC:NOIS ON
Switches on the noise measurement.
Manual operation: See "Noise Measurement State" on page 510
See "Switching All Noise Measurements Off" on page 510
11.8.3.10 Phase Noise Measurement Marker
The following commands control the phase noise measurement marker function.
Analyzing Measurements (Basics)