Remote Commands
R&S
®
FSW
998User Manual 1173.9411.02 ─ 19
MMEMory:SELect:CHANnel[:ITEM]:DEFault
MMEMory:SELect[:ITEM]:DEFault
This command selects the current settings as the only item to store to and load from a
configuration file.
Manual operation: See "Items" on page 548
MMEMory:SELect:CHANnel[:ITEM]:HWSettings <State>
MMEMory:SELect[:ITEM]:HWSettings <State>
This command includes or excludes hardware settings when storing or loading a con-
figuration file.
Hardware settings include:
general instrument configuration
measurement hardware configuration including markers
limit lines
Note that a configuration may include no more than 8 limit lines. This number
includes active limit lines as well as inactive limit lines that were used last.
Therefore the combination of inactivate limit lines depends on the sequence of use
with MMEMory:LOAD:STATe.
color settings
configuration for the hardcopy output
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
MMEM:SEL:HWS ON
Manual operation: See "Items" on page 548
MMEMory:SELect:CHANnel[:ITEM]:LINes:ALL <State>
MMEMory:SELect[:ITEM]:LINes:ALL <State>
This command includes or excludes all limit lines (active and inactive) when storing or
loading a configuration file.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
MMEM:SEL:LIN:ALL ON
Manual operation: See "Items" on page 548