Input of a new Entry ID
Next, you are asked for a new calibration entry ID. If you want to set a
new ID you can enter it here. After verification this new entry ID is
stored internally in the NGMO. Next time you must use this new
number to enter the calibration mode.
If you do not want to change the calibration entry ID, just press the
SHIFT key instead of entering a number.
NOTE:
When using the calibration set-up specified in section 7.2, the
following display may be neglected.
After entering the Entry ID the display should show the following:
Remove all loads and
and connect cal. Adapter
9. Press ENTER.
NOTE:
When using the calibration set-up specified in section 7.2, the
following display may be neglected.
The display must show the following:
on output terminal:A
Connect digital DVM
10. Press ENTER.
The display must show the following:
Look at DVM and ent.
Read value: xx mV
3.00 / 05-2014 7-5