11. Enter the Measuring value.
Use either the variation knob or numerical entry keys (press shift
key to initiate the numerical input) to enter the measured value
that is displayed on the DVM.
12. Press ENTER.
The display must show the following:
Look at DVM and ent.
Read value: xx.yyy V
13. Enter the Measuring value.
Use either the variation knob or numerical entry keys to enter the
measured value that is displayed on the DVM.
14. Press ENTER.
NOTE:
When using the calibration set-up specified in section 7.2, the
following display may be neglected.
The display must show the following:
Connect known 1 Ohm
Resistor to out.term
15. Press ENTER.
The display must show the following:
Enter exact Resistor
value: x.xxxx Ohms
NOTE:
The values of the resistors installed in the calibration adapter
can be taken from the calibration certificate supplied with the
calibration adapter.
7-6 3.00 / 05-2014