16. Enter the resistance value.
Use either the variation knob or numerical entry keys to enter the
exact resistor value.
17. Press ENTER.
NOTE:
When using the calibration set-up specified in section 7.2, the
following display may be neglected.
The display must show the following:
Connect known 33 Ohm
Resistor to out.term
18. Press ENTER.
The display must show the following:
Enter exact Resistor
value: 3x.xxx Ohms
NOTE:
The values of the resistors installed in the calibration adapter
can be taken from the calibration certificate supplied with the
calibration adapter.
19. Enter the resistance value.
Use either the variation knob or numerical entry keys to enter the
exact resistor value.
20. Press ENTER.
NOTE:
When using the calibration set-up specified in section 7.2, the
following display may be neglected.