P
o
w
S
upp
l
y
NG
M
O
Calib
r
ation
The display must show the following:
Conn. Known 3.3 Kohm
Resistor to out.term
21. Press ENTER.
The display must show the following:
Enter exact Resistor
value: 3xxx.x Ohms
NOTE:
The values of the resistors installed in the calibration adapter
can be taken from the calibration certificate supplied with the
calibration adapter.
22. Enter the resistance value.
Use either the variation knob or numerical entry keys to enter the
exact resistor value.
23. Press ENTER.
The display must show the following:
Remove cal. adapter
from output term.:A
24. Press ENTER.
7-8 3.00 / 05-2014