21
Heat 5 units change
increase
The tip of the microelectrode becomes smaller, the
taper gets longer, and the impedance becomes larger
decrease
The tip of the microelectrode becomes larger, the
taper
gets shorter, and the impedance becomes
smaller
Pull 10 units change
increase
The tip of the microelectrode becomes smaller, the
taper gets longer, and the impedance becomes larger
decrease
The tip of the microelectrode becomes larger, the
taper
gets shorter, and the impedance becomes
smaller
Velocity
microelectrodes:
1-5 units change;
Other
microelectrodes:
increase
The tip of the microelectrode becomes smaller, the
taper gets longer, and the impedance becomes larger
decrease
The tip of the microelectrode becomes larger, the
taper
gets shorter, and the impedance becomes
smaller
Time 10 units change
increase
The tip of the microelectrode becomes larger, the
taper
gets shorter, and the impedance becomes
smaller
decrease
The tip of the microelectrode becomes smaller, the
taper gets longer, and the impedance becomes larger
Delay 10 units change
increase
The tip of the microelectrode becomes larger, the
taper
gets shorter, and the impedance becomes
smaller
decrease
The tip of the microelectrode becomes smaller, the
taper gets longer, and the impedance becomes larger
Pressure 100 units change
increase
The tip of the microelectrode becomes smaller, the
taper gets longer, and the impedance becomes larger
decrease
The tip of the microelectrode becomes larger, the
taper
gets shorter, and the impedance becomes
smaller
Pulling the microelectrode under the critical parameter with the cycle value changing
will cause unstable pull results, and above rules may not apply!