37
Drierite can be reused after drying. The specific drying method is as follows:
1) Pour the drierite into a heat-resistant container and spread it out as far as possible.
2) Put the drierite in the oven and dry it at 200 °C for one hour. Observe whether the drierite
turns blue. If not, increase the drying time until it turns blue.
3) After drying, transfer the drierite to the glass container before the drierite temperature
drops.
4) Cool the drierite and then put it back in the canister.
6.5 General check
Please confirm that the filament has been installed and is not damaged;
Please check the shape of the filament before installing. The normal shape of the filament is
shown in Figure 6-20 (right). The deformed heating filament cannot be used. Use tweezers and
other tools to restore the heating filament to its normal shape before use.
Figure 6-20
After replacing the filament, be sure to adjust the new one. Please refer to
6.3 Replace
filament
;
Confirm whether the touch screen operation is normal;
Confirm whether the identification is firmly pasted;
If it is found that some parts or materials need to be replaced during general check or
maintenance, please contact RWD after-sales service for support.