2. Installation
43
Setting the Temperature
1. Turn the Embedding Module power switch on.
NOTE:
The power switch of the Cryo Module may be off. Also note that
Cryo Module temperatures can be set even when the control
cable is not connected.
2. Touch the location corresponding to the temperature you want to set.
3. Touch the "
"/" " buttons to change the set temperature. Touching
the "
" button saves the setting, after which the display returns to the
Operation screen.
Touch the "
" button, and the display will return to the Operation screen
without the setting being saved.
NOTE:
The current temperature displayed on the Operation screen will
blink until it falls within +3/-1˚C of the set temperature.
CAUTION:
Set the temperatures of the heating parts of the instrument
so that the tissues, paraffin, cassettes and base molds can
be kept at the same, appropriate temperature as much as
possible. A large temperature difference prevents the paraffin
block from remaining in a uniform state, and the block may
crack.
CAUTION

Table of Contents

Other manuals for Sakura Tissue-Tek TEC 6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Tissue-Tek TEC 6 and is the answer not in the manual?

Sakura Tissue-Tek TEC 6 Specifications

General IconGeneral
BrandSakura
ModelTissue-Tek TEC 6
CategoryControl Unit
LanguageEnglish