3.2.3 Changing the Parafn Flow Rate
Turning the dial 3.5 times counterclockwise from the fully closed position
(achieved by turning the dial all the way in the clockwise direction), opens the
dial fully.
Turn the paraffin flow rate adjustment dial when paraffin is molten (can be
dispensed).
If the dial is forcibly turned when the parafn is solid, failure may result.
The parafn ow rate for dispensing can be adjusted using the parafn ow
rate adjustment dial.
Continue turning the dial, and it will come off. The dial need not be removed
in normal conditions, but if it was removed, be careful not to lose the spring
inside the dial.
CAUTION:
Do not close the paraffin flow rate adjustment dial of the
Embedding Module fully after the process. If the paraffin
cools down and solidifies when the flow rate adjustment dial
is fully closed, the paraffin may expand the next time it is
melted, and leak out from the flow rate adjustment dial.
パラフィン流量調整ダイヤル
Parafn ow rate adjustment dial
パラフィン流量調整ダイヤル
スプリング(バネ
ワッシャー
Parafn ow rate adjustment dial
Washers
Spring
CAUTION