3. Explanation of Operations
69
3.2.5 Continuing Embedding after the End Time of Auto
Operation
During auto operation, you may want to continue the embedding work
because there are still tissues to be embedded although the end time comes
soon and the instrument enters the paused mode. In this case, you can
continue the embedding process by following the steps below:
1. Touch the “Extend” button [1] on the
Operation screen.
2. The displayed end time temporarily
changes to the next end time [2].
3. Continue the embedding process.
4. When the process is over and you
want to pause the instrument, touch
the “End” button [3].
NOTE:
When the “Extend” button [4] is touched again without touching
the “End” button [3], the instrument will go back to the user-
scheduled auto operation mode. Refer to 2.3.1 Operation
Method for details of the operation of the Extend button.
[1]
[2]
[4] [3]