38
2.1.4 Installation Method
If you have any questions on how to install the instrument, contact the Sakura
instrument distributor or representative.
Before moving the instrument, if it has been previously installed, conrm that
no molten parafn is left inside the Embedding Module or any parafn left in
the module is cool and solid.
CAUTION:
Do not move the instrument while molten paraffin is inside.
Before moving the instrument, remove the power plug and
confirm no molten paraffin inside the instrument or wait for it
to cool and solidify.
Installing the Instrument
1. Select a work bench that has enough space to allow the Embedding
Module and Cryo Module to be placed on, and enough strength to
withstand the mass of the instrument containing parafn (approx. 55 kg).
The left, right and rear side of the instrument can be in contact with walls,
etc. On top of the work bench, at least 80 cm of space is required so that
parafn can be added.
NOTE:
The instrument can be placed on a work bench with a depth of 60
cm or more.
CAUTION
Cryo Module Embedding Module
Approx. 91cm
Approx. 65 cm