5. Troubleshooting
87
Cleaning the Tissue Tray
The parafn inside the tissue tray gradually becomes dirty because this tray
holds cassettes containing tissues. It is recommended to replace the parafn
in this tray with fresh one regularly (after every use).
1. Remove or fully open the lid of the warming chamber, take the tissue tray
out of the warming chamber, and discard the parafn.
CAUTION:
Be careful of handling if the tray is hot.
2. Wash the tissue tray with warm
water (approx. 70 ºC).
3. Wipe down the tray using a lint-
free paper.
Cleaning the Parafn Drain Tray
Discard the parafn in the parafn drain trays every time after use.
Before switching off the Embedding Module, clean the paraffin drain tray
while the parafn is warm and soft.
CAUTION:
Continuing to use the paraffin drain trays without cleaning
may cause them to overflow and dirty the surrounding areas.
1. Remove the parafn on the hot plate. (Refer to Cleaning the Hot Plate
and Cold Spot in this paragraph)
NOTE:
Clean the hot plates first before removing the paraffin drain trays
from the instrument to prevent paraffin residues to drip onto the
bench from the hot plates.
CAUTION
Lint-free paper
CAUTION

Table of Contents

Other manuals for Sakura Tissue-Tek TEC 6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sakura Tissue-Tek TEC 6 and is the answer not in the manual?

Sakura Tissue-Tek TEC 6 Specifications

General IconGeneral
BrandSakura
ModelTissue-Tek TEC 6
CategoryControl Unit
LanguageEnglish