5. When the forceps holder temperature reaches a preset value and the
paraffin surface has slightly softened in each forceps well, a message,
“Heated” is displayed with sounding an alarm again.
NOTE:
A temperature setting to trigger the “Heated” message and the
alarm is user-selectable. Tap the temperature area [4] on the
screen to display the setting window and change to a desired
value.
6. Remove the paraffin from
each forceps well, using
forceps.
7. Wipe down the interior of
each forceps well, using
a lint-free paper or cotton
swab, to remove foreign
matters.
8. After the cleaning is nished,
touch the "
" button [5] to
end the Cleaning mode.
[4]
Solidied parafn
Cotton swab
[5]