5. Troubleshooting
93
5. Wipe down the interior of the parafn chamber using the scraper wrapped
with a lint-free paper, to remove foreign matters.
CAUTION:
When cleaning the instrument, do not use any organic
solvents such as xylene. If any organic solvent flows into the
paraffin line between the paraffin chamber and the dispenser,
the internal parts may degrade or deform, which will cause an
instrument failure.
6. Remove the strainer and wash it with warm water (approx. 70 ºC ).
7. Completely wipe off any water droplets and return the strainer into the
parafn chamber.
Cleaning the wrist rest cushions
Remove parafn attached to the cushions and rinse with warm water at 70°C
or below.
CAUTION
Scraper wrapped with
a lint-free paper
Strainer
Warm water
(approx. 70 ºC)