DRAWING No SP21 18-1805 2040 2040
6-11 End of Measurement
 Be a maximum range of DCV or ACV600 Vif measurement
terminated it.
7MAINTENANCE
7-1 Maintenance and inspection
 1.Appearance
 ・Is the appearance not damaged by falling
 2.Test leads and fuse
 ・Is the cord of the test leads not damaged
 ・Is the core wire not exposed at any place of the test leads
  If your instrument falls in any of the above items, do not use it
and have it repaired or replace it with a new one.
Make sure that the test leads are not cut, referring to the
section 6-1.
7-2 Calibration
 The calibration and inspection may be conducted by the dealer.
For more information, please contact the dealer.
1. This section is very important for safety. Read and understand
the following instruction fully and maintain your instrument
properly.
2. The instrument must be calibrated and inspected at least
once a year to maintain the safety and accuracy.