Entris
®
 BCE Operating Instructions  17
    System Settings 
7.2  Setting the Calibration and Adjustment
7.2.1  Switching the isoCAL Function On or Off 
(Only Model i-1x)
When using the isoCAL function, the device performs an 
automatic time- and temperature-dependent internal 
calibration and adjustment.
If this relates to a conformity-assessed device in 
legal metrology: In some cases it is not possible to 
switch off the isoCAL function. 
Procedure
 t Open the “SETUP”/”BALANCE” menu.
 t To set the automatic start of the isoCAL function: Select the 
“ON” calibration value for the “ISOCAL” parameter.
 t To set the manual start of the isoCAL function: Select the 
“Note” calibration value for the “ISOCAL” parameter.
 t To switch off the isoCAL function: Select the “OFF” 
calibration value for the “ISOCAL” parameter.
7.2.2  Setting Internal Calibration and Adjustment 
(Only Model i-1x)
The following functions can be set for the internal calibration 
and adjustment:
 − Internal calibration with automatic start of the adjustment.
 − Internal calibration with manual start of the adjustment.
Procedure
 t Open the “SETUP”/”BALANCE” menu. Call up the “CAL.JUST.” 
parameter and select the value “CAL.INT.”.
 t If the calibration function needs to be set with subsequent 
automatic adjustment: In the “SETUP”/”BALANCE” menu, for 
the “CAL.SEQ.” parameter, select the “ADJUST” setting value.
 t If the calibration function needs to be set without 
subsequent automatic adjustment: In the 
“SETUP”/”BALANCE” menu, for the “CAL.SEQ.” parameter, 
select the “CAL./ADJ.” setting value.
7.2.3  Setting the External Calibration and Adjustment
The following functions can be set for the external calibration 
and adjustment:
 − External calibration with manual start of the adjustment.
If this relates to a conformity-assessed device in 
legal metrology: External calibration and adjust-
ment is not possible.
Procedure
 t Open the “SETUP”/”BALANCE” menu. 
 t If the calibration function needs to be set with subsequent 
automatic adjustment: Select the “ADJUST” calibration 
value for the “CAL.SEQ.” parameter.
 t If the calibration function needs to be set without 
subsequent automatic adjustment: Select the “CAL-ADJUST” 
calibration value for the “CAL.SEQ.” parameter. 
Setting the Weight Value for the External Weight
A preset weight value or a user-defined weight value can be set 
for the external weight. 
Procedure
 t If the preset weight value needs to be used: In the 
“SETUP”/”BALANCE” menu, for the “CAL./ADJ.” parameter, 
select the “EXT.CAL.” setting value.
 t If a user-defined weight value needs to be set: 
 t In the “INPUT” menu, select the “CAL.WT.” setting value.
 t Enter the desired weight value and press the [Confirm] 
button.
 t To use the user-defined weight value for the next 
calibration: In the “SETUP”/”BALANCE” menu, for the 
“CAL./ADJ.” parameter, select the “E.CAL.USR” setting 
value.