38 Entris
®
BCE Operating Instructions
Technical Data
15.10 Metrological Data
15.10.1 Models BCE6202 | BCE4202 | BCE3202
Model
BCE6202-1x |
BCE6202i-1x
BCE4202-1x |
BCE4202i-1x
BCE3202-1x |
BCE3202i-1x
Unit Value Value Value
Readability | Scale interval (d) mg 10 10 10
Maximum capacity (Max) g 6200 4200 3200
Repeatability
At 5%, typical value mg 5 5 5
At approx. the maximum load, typical value mg 10 10 10
Linearity deviation
Limits ± mg 20 20 20
Typical value ± mg 6 6 6
Sensitivity drift between +10 °C and +30 °C ± ppm/K 2 2 2
Tare maximum capacity (subtractive) <100% of maximum capacity
isoCAL (only for i-1x models):
Temperature change K 2 2 2
Time interval h 6 6 6
Only for models with approval:
Accuracy class II II II
Type BC_EE BC_EE BC_EE
Verification scale interval (e) mg 100 100 100
Minimum load (Min) mg 500 500 500
Minimum initial weighing according to USP (United States
Pharmacopeia), Chap. 41
Optimum minimum initial weighing g 8.2 8.2 8.2
Typical measurement time s
<1.0 <1.0 <1.0
Typical stabilization time s
<0.9 <0.9 <0.9
Recommended Calibration Weight
External calibrated test weight g 5000 2000 2000
Accuracy class in accordance with OIML R111-1 F1 F1 F1
Weighing pan size mm 182 x 182 182 x 182 182 x 182
Net weight, approx. kg 4.6 | 6.2 4.6 | 6.2 4.6 | 6.2