7.3  Parameter List
7.3.1  “SETUP”/”BALANCE” Menu
Parameter Setting values Explanation
AMBIENT V.STABLE Sets the ambient conditions to “very stable”: Activates a fast change in the weight values in 
the event of a load change with a high output rate.
Recommended for the following work environment:
 − Very stable table near the wall
 − Closed and calm room
STABLE* Sets the ambient conditions to “stable”. Recommended for the following work environment:
 − Stable table
 − Slight movement in the room
 − Slight draft
UNSTABL. Sets the ambient conditions to “unstable”: Activates the delayed change in weight values with 
a reduced output rate. Recommended for the following work environment: 
 − Simple office desk
 − Room with moving machinery or personnel
 − Slight air movement
V.UNSTBL. Sets the ambient conditions to “very unstable”: Activates a significantly delayed change in the 
weight values and long wait for stability with a further reduction in the output rate.
Recommended for the following work environment:
 − Noticeable and slow floor vibrations
 − Noticeable building vibrations
 − Weighed goods moved
 − Very strong air movements
APP FILT. FINAL.RD.* Activates a filter that enables a fast change in the display for very fast load changes. 
Display changes with minimal load changes (in the digit range) occur more slowly.
FILLING Activates a filter that enables a very fast change in the display with minimal load changes, 
e.g. when filling containers.
STABIL. V. ACC. Sets the stability to “very accurate”.
ACC.* Sets the stability to “accurate”.
FAST Sets the stability to “fast”.
ZER./TAR. W/O STB. If the button is pressed: The function of the [Zero] or [Tare] button is executed immediately.
W/ STAB.* The function of the [Zero] or [Tare] button is only executed after stability is achieved.
AUTOZER. ON* Activates automatic zeroing. The display is automatically set to zero in case of a deviation of 
0 less than (X).
OFF Deactivates automatic zeroing. Zeroing must be triggered with the [Zero] button. 
* Factory setting