42 Entris
®
BCE Operating Instructions
Technical Data
15.10.5 Models BCE8200 | BCE6200
Model
BCE8200i-1x
BCE8200-1x
BCE6200i-1x
BCE6200-1x
Unit Value Value Value Value
Readability | Scale interval (d) mg 1000 1000 1000 1000
Maximum capacity (Max) g 8200 8200 6200 6200
Repeatability
At 5% load, typical value mg 500 500 500 500
At approx. the maximum load, typical value mg 1000 1000 1000 1000
Linearity deviation
Limits ± mg 1000 1000 1000 1000
Typical value ± mg 600 600 600 600
Sensitivity drift between +10 °C and +30 °C ± ppm/K 2 7 2 7
Tare maximum capacity (subtractive) <100% of maximum capacity
isoCAL (only for i-1x models):
Temperature change K 2 2
Time interval h 6 6
Only for models with approval:
Accuracy class II II II II
Type BC_EG BC_EI BC_EG BC_EI
Verification scale interval (e) mg 1000 1000 1000 1000
Minimum load (Min) mg 50000 50000 50000 50000
Minimum initial weighing according to USP
(United States Pharmacopeia), Chap. 41
Optimum minimum initial weighing g 820 820 820 820
Typical measurement time s
<1.0 <1.0 <1.0 <1.0
Typical stabilization time s
<0.9 <0.9 <0.9 <0.9
Recommended Calibration Weight
External calibrated test weight g 5000 5000 5000 5000
Accuracy class in accordance with OIML R111-1 F2 F2 F2 F2
Weighing pan size mm 182 x 182 182 x 182 182 x 182 182 x 182
Net weight, approx. kg 6.2 4.3 6.2 4.3