Entris
®
BCE Operating Instructions 41
Technical Data
15.10.4 Models BCE8201 | BCE5201 | BCE2201
Model
BCE8201i-1x
BCE8201-1x
BCE5201-1x (x =
only CCN, OIN) |
BCE5201i-1x
BCE5201-1x
BCE2201-1x (x =
only CCN, OIN) |
BCE2201i-1x
BCE2201-1x (x =
only S, SAR, SJP,
SKR)
Unit Value Value Value Value Value Value
Readability | Scale interval (d) mg 100 100 100 100 100 100
Maximum capacity (Max) g 8200 8200 5200 5200 2200 2200
Repeatability
At 5% load, typical value mg 50 50 50 50 50 50
At approx. the maximum load, typical value mg 100 100 100 100 100 100
Linearity deviation
Limits ± mg 100 300 100 300 100 300
Typical value ± mg 60 100 60 100 60 100
Sensitivity drift between +10 °C and +30 °C ± ppm/K 2 7 2 7 2 7
Tare maximum capacity (subtractive) <100% of maximum capacity
isoCAL (only for i-1x models):
Temperature change K 2 2 2
Time interval h 6 6 6
Only for models with approval:
Accuracy class II II II II
Type BC_EG BC_EI BC_EE BC_EE
Verification scale interval (e) mg 1000 1000 100 100
Minimum load (Min) mg 5000 5000 5000 5000
Minimum initial weighing according to USP
(United States Pharmacopeia), Chap. 41
Optimum minimum initial weighing g 82 82 82 82 82 82
Typical measurement time s
<1.0 <1.0 <1.0 <1.0 <1.0 <1.0
Typical stabilization time s
<0.9 <0.9 <0.9 <0.9 <0.9 <0.9
Recommended Calibration Weight
External calibrated test weight g 5000 5000 5000 5000 2000 2000
Accuracy class in accordance with OIML R111-1 F2 F2 F2 F2 F2 F2
Weighing pan size mm 182 x
182
182 x
182
182 x
182
182 x
182
182 x
182
182 x
182
Net weight, approx. kg 6.2 4.3 4.3 | 6.2 4.3 4.3 | 6.2 4.3