22 Entris
®
BCE Operating Instructions
System Settings
Parameter Setting values Explanation
LIGHT
OFF Deactivates the lighting on the operating display.
ON* Activates the lighting on the operating display.
* Factory setting
7.3.6 “DATA.OUT.”/”COM. SBI” Menu
Parameter Setting values Explanations
COM. OUTP. IND.NO* Activates the manual data output without stability.
IND.AFTR Activates the manual data output after stability.
AUTO.W/O Activates the automatic data output without stability.
AUTO W/ Activates the automatic data output after stability.
STOP.AUT. OFF* Deactivates the option to abort the automatic data output.
ON The automatic data output is aborted by pressing the [Print] button or a software command.
AUTO.CYCL. EACH VAL* Starts the automatic data output with a cycle after each value.
AFTER 2 Starts the automatic data output with a cycle after every 2nd value.
INTERV. The output rate can be set from 0 - 9999 seconds under “INPUT/INTERV.”.
FORMAT 22 CHARS* The data output provides 22 characters per line (16 characters for the measured value and
6 characters for identifiers).
16 CHARS The data output provides 16 characters per line for the measured value.
EXTR.LIN. The data output provides an additional line with the date, time, and weight value.
AUTO.TAR. OFF* Deactivates automatic taring after data output.
ON The device automatically tares after data output.
* Factory setting
7.3.7 “DATA.OUT.”/”PRNT.PAR.” Menu
Parameter Setting values Explanation
ACTIVAT. MAN. NO Manual without stability: The print process can be started manually at any time.
MAN.AFTR* Manual after stability: After pressing the [Print] button, the print command is only executed
once stability is achieved.
INTERV. The output rate can be set from 0 - 9999 seconds under “INPUT/INTERV.”.
AUTO.LC Automatically at load change: The print process starts after every load change.
FORMAT 22 CHARS* The printer output prints 22 characters per line (16 characters for the measured values and
6 characters for identifiers)
EXTR.LIN. The printer output prints an additional line with the date, time, and weight value.
PRT.INIT. OFF Deactivates the output of the application parameters.
ALL* The print command prints all parameters.
MAIN.PAR. The print command only prints the main parameters.
GLP OFF* Deactivates the GLP printout.
CAL./ADJ. Activates the GLP printout for all calibration and adjustment processes.
ALWAYS The GLP printout is always switched on. All printouts contain a GLP header and a GLP footer.
TAR./PRT. OFF* Deactivates automatic taring after printer output.
ON Automatically tares the device after every printout.
* Factory setting