EasyManua.ls Logo

Sartorius FC - Page 112

Sartorius FC
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cabling Diagram
112
Verification of the
Intrinsic Safety
FC/FCA/IS......-.X.... /
Rev.
00
Klausgrete
Weitemeier
Klausgrete
Name
12.02.02
12.02.02
12.02.02
Sheet
of
2
4
35520-099-61-A4
EX
Released by
Reviewed by
Written by
Date
RS-232 data output with control lines
Data output
1)
Uo 8,6 V
Io 23 mA
Po 50 mW
Co 6 µF
Lo 60 mH
Compact scale
FC......-.X....
FCA......-.X....
Weighing platform
IS......-.X....
II 2 G EEx ib IIC T4
II 1 D T135°C (optional)
IIC T4
T-connektor
2)
YTE02-X
Foot
actuated
switch
2)
YPE05-X
Barrier
YDI02-Z
YDI03-Z
II (2) G
[EEx ib] IIC
Cable: install as stationary
and protected equipment !
Maximum cable length 70m
for a 10-wire standard cable
with a maximum of 200 nF/
km and 1mH/km.
The length of the installed
cable is defined as 20 m
because of the RS-232
specifications.
Circuit: RS-232
1)
Control signals (only for YDI03-Z)
1)
Uo 12 V 12 V
Io 82 mA 7.5 mA
Po 240 mW 22 mW
Co 1.41 µF 1.41 µF
Lo 5.5 mH 5.5 mH
Lo/Ro 144µH/ 1.5 mH/
Data output RS-232 Control signals
(pins A/J/K/N and M)
1)
(pins C/D/E/F/G and M)
1)
Ui 12.6 V 12.6V
Ii 85 mA 10 mA
Pi 270 mW 30 mW
Ci 3 nF 4 nF
Li 0 0
1: per circuit
2: optionally used (passive wiring or switch only)
System appraisal
(see EC-Type Examination Certificate KEMA 02ATEX1108 X):
Co <= 125nF; Lo <= 100µH; Lo/Ro <= 33µH/
Non hazardous area
Hazardous area