Calibration/Adjustment
38
Preparation
Set the parameters for calibration and adjustment; e.g., with manual calibration/adjustment, isoCAL off
Step Press key(s) Display/Output
(or follow instructions)
1. Switch on the scale, e Sartorius logo and self-test
if not already on
2. Select the Setup menu H
3. Select “Balance/scale functions”
O soft key
4. Select “Calibration/adjustment”
O soft key
5. Select CAL/isoTST key function
O soft key
o = last setting selected
6. Select desired function and
Q soft key, repeatedly,
confirm (e.g., “Ext. cal./adj.; if necessary
l soft key
factory-def. wt.”)
7. Exit CAL/isoTST key function
o soft key
8. Select “Cal./adjustment sequence”
q soft key