EasyManua.ls Logo

Sartorius FC - Totalizing; Purpose; Features

Sartorius FC
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Totalizing
L
Purpose
This application program runs as
a cumulative memory function.
You can use this application in
combination with any program chosen
from Application 1 (such as counting,
weighing in percent) and one from
Application 2 (checkweighing,
time-controlled functions).
Features
Totalization of weight values and
calculated values
Optional configuration in the Setup
menu for simultaneous storage of net
and calculated values
Optional configuration in the Setup
menu for loading weight values
and calculated values either from
Application 1 (e.g., counting, weighing
in percent) or from Application 2
(checkweighing, time-controlled
functions)
Totalization memory for up to
65,535 values
Simultaneous display in the text line
of transaction counter and, e.g., the
current total
Optional configuration in the Setup
menu for having the scale tare auto-
matically after a value is stored in the
totalization memory, if no preset tare
has been entered
Manual input of the number of
individual weighing operations and
confirmation using the
nDef soft key
(target no. of operation nDef). Result
printed and memory cleared after
printout of nDef.
Optional configuration in the Setup
menu to add the current weight,
with display accuracy, to the current
total by pressing the
M+ soft key
and generate a printout of the result
Optional configuration in the Setup
menu for stability-dependent storage
of the measured value: Balance/
scale functions,
Stability range
Optional automatic storage of
measured values
Storage of measured value is indicated
by
Uu; zz indicates that you can
place a load on the scale.
Minimum load threshold for automatic
storage
Press the
M- soft key to delete the last
value added to the totalizing memory.
The transaction counter value is
reduced by one and a printout is
generated.
Press the
MR soft key for information
about number of transactions and the
current total. In the Setup menu, you
can define whether the information is
displayed and printed, or only printed,
and whether the information comprises
an intermediate or final evaluation
In the Info window you can choose
which value is displayed in the text line
during weighing
Printout of the end result independent
of which program is configured for
Application 1 or Application 2.
Configure the Setup menu to define
which values are included on the
printout (printout of individual
components)
Press the key identified by
MR (soft key
label) for a printout of an intermediate
evaluation after each addition or a final
evaluation
If you end the totalization process by
pressing c without having first
pressed the
MR soft key for a printout,
a final evaluation is printed when you
press c
Optional configuration in the Setup
menu to clear the totalizing memory
and reset the transaction counter by
pressing c or after an evaluation is
printed out
Totalization data and transaction
counter data are stored in non-volatile
memory
Continue totalization after turning the
scale off and back on
Factory Settings of Parameters
Automatic storage:
Off
Minimum load for automatic storage:
10 digits
Source of data for auto storage:
Application 1
Evaluated values: Net
Evaluation mode, MR key function:
Intermediate
evaluation, print
M+/M– function, then tare: Off
Printout of individual components: Yes
Stability range: 2 digits
Application-defined output:
Print then tare:
Off
Soft Key Functions
M+ Add weight values or
application values to the total
in the totalization memory.
The component or transaction
counter value increases by one
each time you press this key.
M- Delete the last value added
to memory. The transaction
counter value decreases by 1.
You cannot delete previous
values by repeatedly pressing
this key.
MR Print or display an inter-
mediate or final evaluation
nDef Store the input number
of components
Printout for Totalizing
The transaction or component counter is
printed in front of each measured value
(weight). When an intermediate or final
evaluation is printed out, all results up to
this point are included.
n5
Total +151.67 g
72