58
Purpose
With this application program you can
compensate for over-poured components
in formulation.
If a component is over-poured when
weighing in the individual formulation
components, the mixture already poured
cannot be used in its current composition.
To avoid having to discard the materials
weighed, you can adjust the proportions
of the formulation to compensate for the
over-pour.
When you use this application, the
recalculation procedure is mainly
performed by the scale.
You can use this application program
in combination with a program chosen
from Application 2 (checkweighing,
time-controlled functions) and one from
Application 3 (totalizing, formulation,
statistics).
Available Features
Individual components (up to 99)
weighed in with a readout showing from
“0” to the desired component weight
Transaction counter shows the next
component expected
Weighed components are stored,
followed by automatic printout
and taring
Additive weighing of components
with printout
Toggle the display between component
weight and total formulation weight
(additive mode) after first component
is stored
Stored component weight displayed
as true net weight for 2 seconds
Enter a divisor before or during
component weighing. For example, if
the formulation has a total weight of
100 g, enter the divisor 10 to weigh
in a total formulation of 1,000 g
If a component is over-poured, you
can use the recalculation function to
change the amount of this component
indicated in the formulation by using
plus or minus keys or numeric input.
A factor is then calculated by which all
components amounts will then adjusted
Recalculation factor displayed in the
text line, with a warning symbol if
the factor is not equal to 1
All components displayed with number
and the amount (by weight) to be
added in follow-on filling. Components
displayed in sequence by the scale
Display of actual net weight during
follow-on filling
After the amounts of the components
already weighed have been corrected,
weighing continues according to
the adjusted formulation amount.
The readout is recalculated (updated)
according to the divisor
You can repeat the over-pour correction
procedure as often as necessary, in case
other components are over-poured
After follow-on (corrective) filling, the
total amount differs from that given for
the formulation, but the proportion of
components in relation to each other is
the same
You can have the weight printed after
each measurement
Choose whether the current component
weight or tare value is printed after each
measurement
Individual component weights are
printed as “Compxx.”
Press c to exit the application
program. The component memory is
cleared and the sum of components
printed as “S-Comp.”
Toggle between the recalculation
program and other applications
(e.g., checkweighing) by pressing D.