“FlexPrint” Printout Function
93
Purpose
The YAD02IS “Nice Label Express”
software from Sartorius lets you load
user-defined label printing formats and
the corresponding print instruction files
in your scale. This software lets you
connect any of a number of printers,
equipped with a variety of printer fonts,
to the RS-232 interface on your scale.
Features
With the “FlexPrint” option activated:
Print command generates configured
printout (if print instruction file exists;
see table, next page). Print command
generates default printout (if print
instruction file does not exist).
The function that generates an
automatic printout upon initialization
of an application cannot be used.
Initialization data can be output
only to a print instruction file.
The following items are output only
as standard printouts:
Calibration/adjustment
SETUP printouts
p key in “Identifier:” display page
p key on “Product data memory:”
display page
p on “Parameter” display page
for “Checkweighing” application:
print function carried out
MR function carried out when p is
pressed during evaluation (info window)
in Totalizing and Statistics applications.
To recall the file names, software ID
and version numbers, for FlexPrint, see
the section on “Basic Settings,” Info
Display.”
Printout for legal metrology:
Weight blocks (special weight value
formats that are acceptable in legal
metrology) are designated by the
following line, which is printed in
both header and footer and cannot
be edited:
------ CE [M] ------
Examples of Weight Block Printouts
Without tare:
------ CE [M] -------
N + 348.65 kg
------ CE [M] -------
With tare:
------ CE [M] -------
G + 459.70 kg
N + 348.65 kg
T1 + 111.05 kg
------ CE [M] -------
Scale tare (2nd tare memory):
------ CE [M] ------
G + 124.45 kg
N + 100.00 kg
T1 + 24.00 kg
T2 + 0.45 kg
------ CE [M] ------
Preparation
§ Turn on the scale: press the e key
§ Configure FlexPrint in Setup:
press the H key
§ Select
Printout: press q soft key
3 times, and then
O soft key
§ Confirm
Application-
defined output:
press
O soft key
§ Select
FlexPrint: press q soft
key 3 times and then
O soft key
§ Select
On: press q soft key and
then
O soft key
§ Save settings and exit Setup:
press
oo soft key
Print events
Printouts generated using the “Nice Label Express” software are divided into the following groups:
Event Explanation File name for event group:
1. p key with individual values Print key PPRINT
2. p key with text input Input and p key PDIRECT
3. GLP /GMP header GLP header PGMPHEAD
4. GLP /GMP footer GLP footer PGMPFOOT
5. Results, Application 1 Animal weighing, MR-CF PA1RES
6. Results, Application 2 OK values, time-controlled print PA2RES
7. Results, Application 3 MR, MR-CF PA3RES
8. Components, Application 1 M+ printout PA1COMP
9. Components, Application 3 M+/M– printout PA3COMP