54
Purpose
Use this program to determine the
weights of unstable samples (e.g., live
animals) or to determine weights under
unstable ambient conditions. In this
program, the scale calculates the weight
as the average of a defined number of
individual weighing operations. These
weighing operations are also known as
“subweighing operations.”
You can use this application program
in combination with a program chosen
from Application 2 (checkweighing,
time-controlled functions) and one from
Application 3 (totalizing, formulation,
statistics).
Available Features
Animal weighing started manually or
automatically
Automatic start:
when a defined threshold has been
exceeded (Minimum load threshold:
None; 10; 20; ...; 500; 1,000 display
increments)
when three successive subweights lie
within a user-defined tolerance range
(calm; normal; active; 0.1%; 0.2%; ...;
50%; 100% of the animal/object)
Manual start:
also possible when the load is under
the minimum load threshold
when three successive subweights lie
within a user-defined tolerance range
(calm; normal; active; 0.1.%; 0.2%;
...; 50%; 100% of the animal/object)
Optional scale configuration in
the Setup menu for automatically
initializing this application when
you turn on the scale (Setup:
Printout: Auto-start application
when power goes on: On)
Number of weighing operations for
calculation of an average (
mDef) can
be set before the beginning of each
animal weighing operation
The factor for calculation of the result
can be set before the beginning of each
animal weighing operation
The number of subweighs remaining
to be performed is indicated in the
text display during weighing
Arithmetic average displayed as
a result in the pre-set weight unit
(identified by the a symbol).
Optional multiplication of the
arithmetic average by a user-defined
factor
Mul.
A circle o” is displayed as weight unit
and
Mul = xxx is shown in the
text line
Toggling between the weight and the
calculated value by pressing the
xNet
soft key and the xRes soft key
Automatic output of results via the
interface port:
– Number of weighing operations
mDef
– Multiplication factor Mul
Automatic output of results (printout)
via the interface port:
– Weighing result
xNet
– Calculated result xRes
The following options have to be set:
Printout: Application-defined output:
Auto print upon initialization: All values
The unload threshold is equal to
one-half the minimum scale capacity
Return to weighing mode by unloading
the scale; i.e., when the load is below
the unload threshold
Factory Settings
Animal activity:
5% of the animal/object
Start: Automatic
Minimum load for automatic storage:
100 display increments
Decimal places in result display:
2 decimal places
Printout:
Average weight only
Soft Key Functions
New Automatic start:
Unload scale and weigh
next animal, if desired
Press key to start next
subweigh
Manual start:
Start next subweigh
mDef Store user-defined number
of subweighs for averaging
Mul Store user-defined factor
as multiplication factor for
calculated the arithmetic
mean
xNet Toggle to the animal weighing
application
xRes Toggle to the calculated
animal weighing result
Start Activate Animal weighing
Printout for Animal Weighing
Upon completion of the averaging process,
you can have the results printed out
automatically. You can also have both the
weight and the calculated result printed.
mDef 10
Mul 0.00347
xNet + 153.00 g
xRes + 5.30 o
mDef: Number of subweighing
operations for averaging
Mul: Multiplication factor
xNet: Result of averaging
xRes: Calculated result
Animal Weighing