Loading...
Cabling Diagram
114
Verification of the
Intrinsic Safety
FC/FCA/IS......-.X.... /
Rev.
00
Klausgrete
Weitemeier
Klausgrete
Name
12.02.02
12.02.02
12.02.02
Sheet
of
4
4
35520-099-61-A4
EX
Released by
Reviewed by
Written by
Date
10mA data output
Converter
725324
[EEx ib] IIB
IIC T4
In accordance with the EC-Type Examination Certificate
KEMA 01ATEX1099X the converter 725324 may be
connected to the FCT01-X.., but with reduced parameters:
Co = 620nF
Lo = 2 mH.
Cable: install as stationary and
protected equipment. Maximum
length 600m for a standard
cable with a maximum of
200nF/km and 1mH/km.
Non hazardous area
Hazardous area
Compact scale
FC......-.X....
FCA......-.X....
Weighing platform
IS......-.X....
II 2 G EEx ib IIC T4
II 1 D T135°C (optional)
TTY Data output
1)
(pins G/K and D/F/J) (pins C/E and D/F/J)
Ui 14.7 V 14.7 V
Ii 50 mA 130 mA (resistively limited)
Pi 265 mW 100 mW
Ci 0 0
Li 0 0
1: per circuit