Selecting the Printout Function (PRINTOUT)
Printout Parameters (Overview)
o factory setting
user-defined setting
Setup Application- Stability parameter Without stability
1)
Print- defined ο With stability
output
Print on request, ο Off
then tare On
Auto print upon ο Off
initialization All values
(of the application) Only main values
Configured printout See own chapter starting on the next page
FlexPrint ο Off
On
Automatic Stability parameter ο Without stability
output of At stability
displayed value
3)
Automatic after weight change
2)
Stop auto print Use print key p
ο Not possible
Time-dependent ο 1 display update
auto print 2 display updates
10 display updates
4)
100 display updates
4)
Output to Serial communication ο Application-defined output
interface port (PERIPHERALS) Automatic output of displayed value
Line format For raw data (16 characters)
ο For other apps/GLP (22 characters)
ISO/GLP/GMP ο Off
printout Only for calibration/adjustment
Always on
Identification Lot ID (L ID): Enter 20 characters max
ID1: ID1
ID2: ID2
ID3: ID3
ID4: ID4
Factory settings: No
only printout Yes
1
) = When the scale is used in legal metrology, this setting may be configured for control purposes only; printout of values is not permitted
2
) = Autoprint when load change is >10 d and stability is reached: no printout until residual difference in load value >5 d
3
) = To enable automatic output of the displayed value, the corresponding option (“Automatic output of displayed value”) must be activated
under the next menu item at this level, “Output to interface port”
4
) = These settings may not be changed on verified scales