Printout Configuration 
Practical Example:
Configure an Individual Printout for Counting Application to Include Dotted Line, Date/Time, Piece Count and Net Weight
Settings (changes in the factory settings required for this example):
Setup: Application parameters: Application 1: Counting
Exit the Setup menu: press the 
oo soft key
Then call Setup again and select: Printout: Application-defined output: Configured printout
Step Press key(s)  Display/Output
(or follow instructions)
1. Select Setup menu,  H, then
“Printout” 
q repeatedly and
O soft key
2. Confirm “Application-defined
O soft key
output”
3. Select and confirm
q soft key 3+
“Configured printout”
O soft key
4. Confirm “Indiv. printout”
O soft key
5. Select “Dotted line”
O, q, l soft keys
6. Select “Date/time”
q soft key twice,
then l soft key
7. Select “Piece count”
q repeatedly, 
then l soft key
8. Select “Net weight”
Q soft key repeatedly, 
then l soft key
9. Exit “Printout configuration”
oo soft key
10. Perform weighing operations, p
--------------------
then print 14.01.2000 09:19
Qnt + 598 pcs
N + 2003.13 g
26