Calibration/Adjustment 
Internal Calibration/Adjustment
First set either 
Internal
cal./adjustment or
Selection mode (factory 
setting) in the Setup: Scale menu.
Inside the scale housing is a built-in
motorized calibration weight.
The internal calibration/adjustment
sequence is as follows:
§ Select the calibration function: 
Press the 
isoTST soft key twice
> The internal calibration weight 
is applied automatically
> The scale is calibrated
> If the setting 
Calibrate,
then auto adjust is selected
in the Scale menu, the scale is now
automatically adjusted
> If the setting 
Calibrate,
then manual adjust
is selected in the Scale menu, the 
internal calibration routine is now
ended without adjusting the scale 
(see “Calibration and Adjustment
Sequence”, next column)
> The internal calibration weight is
removed
> (ISO/GMP printout: see page 100)
Calibration and Adjustment Sequence
In the Setup menu, you can configure
the scale so that:
– calibration is always followed 
automatically by adjustment 
Calibrate, then auto
adjust (factory setting) or
– you have the choice of ending the
sequence or starting adjustment 
after calibration 
Calibrate,
then manual adjust
If no deviation is determined in
calibration, or the deviation is within
the tolerance limits dictated by the
degree of accuracy you require, it is 
not necessary to adjust the scale. In 
this case, you can end the calibration/
adjustment sequence after calibration. 
There are 2 soft keys active at this
point:
–
Start to start adjustment
–
End to end the sequence