Calibration/Adjustment
Internal Calibration/Adjustment
First set either
Internal
cal./adjustment or
Selection mode (factory
setting) in the Setup: Scale menu.
Inside the scale housing is a built-in
motorized calibration weight.
The internal calibration/adjustment
sequence is as follows:
§ Select the calibration function:
Press the
isoTST soft key twice
> The internal calibration weight
is applied automatically
> The scale is calibrated
> If the setting
Calibrate,
then auto adjust is selected
in the Scale menu, the scale is now
automatically adjusted
> If the setting
Calibrate,
then manual adjust
is selected in the Scale menu, the
internal calibration routine is now
ended without adjusting the scale
(see “Calibration and Adjustment
Sequence”, next column)
> The internal calibration weight is
removed
> (ISO/GMP printout: see page 100)
Calibration and Adjustment Sequence
In the Setup menu, you can configure
the scale so that:
calibration is always followed
automatically by adjustment
Calibrate, then auto
adjust (factory setting) or
you have the choice of ending the
sequence or starting adjustment
after calibration
Calibrate,
then manual adjust
If no deviation is determined in
calibration, or the deviation is within
the tolerance limits dictated by the
degree of accuracy you require, it is
not necessary to adjust the scale. In
this case, you can end the calibration/
adjustment sequence after calibration.
There are 2 soft keys active at this
point:
Start to start adjustment
End to end the sequence