Calibration/Adjustment 
42
External Calibration/Adjustment* with 
a User-Defined Calibration Weight
First set either 
Ext. cal./
adj.;user-defined wt. or
Selection mode (factory setting)
in the Setup: Scale menu.
You can define a weight for calibration/
adjustment. External calibration/
adjustment must be performed with
weights that are traceable to a national
standard and that have error limits which
are at least 1/3 of the required tolerance 
of the display accuracy. The defined
weight must equal at least 10% of the
maximum scale capacity.
See page 41 for the external calibration/
adjustment sequence. For this example,
select external calibration/adjustment with
a user-defined weight.
The scale has a factory-set weight value
(see “Specifications”).
To reset a user-defined calibration
weight to the original factory setting:
$ Enter the factory-defined value 
manually (see “Specifications”) 
Define the Calibration Weight
Step Press key(s)  Display/Output
(or follow instructions)
1. Select Setup menu H
2. Select “Balance/scale functions”
O soft key
3. Select “Calibration/adjustment”
O soft key
4. Select parameter for external
q soft key 5 x
weight
O soft key
5. Select “Cal./adj. wt.”
q soft key
3000 = last setting selected
6. Enter calibration weight 5 0 0 0
(e.g., 5000 g)
and save
7. Save the calibration weight
l soft key
8. Exit the Setup menu
oo soft key