Calibration/Adjustment
Printout of Calibration/Adjustment
Data
You can have the results of a
calibration/
adjustment procedure printed out.
You can configure whether the printout
is generated as soon as the procedure
is completed, or whether a number of
calibration/ adjustment procedures (up
to 50) are collected for a block printout.
Block Printout of Calibration/
Adjustment Results
With the following menu configuration
under Setup: Balance/scale functions:
Calibration/adjustment, you can store
the data from up to 50 calibration/
adjustment procedures and have them
printed on request:
isoCAL printout
On request, from
record memory
When the memory contains 50 data
records:
additional records are output
automatically
If at least one block printout data
record has been configured, the
following soft keys are available after
you press the
Cal soft key:
Info The number of records is
displayed in the text line
PrtPro Print accumulated records
DelPro
Delete accumulated
records; records can only be
deleted after a printout has
been generated.
If a password has been
assigned in the Setup:
Input menu, you must enter
either the configured pass-
word or the General Pass-
word before you can delete
the records.
For internal calibration/adjustment,
the initialization mode of the procedure
is displayed in the
Start line.
--------------------
13.05.1997 09:17
SARTORIUS
Mod. FC6CCE-HX
Ser. no. 60419914
Ver. no. 01-35-18
ID
--------------------
24.04.1997 12:03
Start: manual
Diff. + 0.01 g
External calibration
completed
25.04.1997 12:10
Start: isoCAL/temp
Diff. + 0.01 g
Internal adjustment
completed
Diff. + 0.00 g
25.04.1997 18:30
Start: Adj.time
Diff. + 0.01 g
Internal adjustment
completed
Diff. + 0.00 g
26.04.1997 9:37
Start: manual
Diff. + 0.01 g
Internal adjustment
completed
Diff. + 0.00 g
27.04.1997 11:53
Start: Ext.cal.
WID
Nom. + 2000.00 g
Diff. + 0.01 g
External calibration
completed
Diff. + 0.00 g
--------------------
13.05.1997 09:17
Name:
--------------------
GLP header
List of Calibration/Adjustment Procedures:
Example 1:
External calibration
Example 2:
isoCAL triggered by difference
in temperature
Example 3:
isoCAL at defined time
Example 4:
Internal calibration/adjustment
triggered manually
Example 5:
External calibration/adjustment
GLP footer
44