Additional Functions
In addition to the functions for:
– alphanumeric input,
– taring (not during
alphanumeric input),
– printing,
you can also access the following
functions from this application:
Calibration/Adjustment
§ Press the
isoTST soft key
> See “Calibration/Adjustment” for
further instructions
Toggling to the Next Application
§ Press D
> See the section on the corresponding
application program for further
instructions
Setup (setting parameters)
§ Press H
> See “Configuration” for further
instructions
Turning Off the Scale
§ Press e
> The scale shuts off
49
Preparation
To calculate a piece count, the average weight of one piece must be known. This average
piece weight can be entered into the Counting program in one of three ways:
– Enter the average piece weight using the numeric keys and store it;
– The last reference sample quantity entered is loaded and displayed when you turn on the
scale. Place the same number of parts on the scale and initialize the Counting program;
– When the automatic initialization parameter (see previous page) is on
(Setup: Printout: Application-defined output: Auto print upon initialization:
All values), the scale goes into the “Counting” mode when you turn it on and loads
the last average piece weight and corresponding reference sample quantity that
were entered or calculated.
Reference Sample Updating
You can have the average piece weight updated during counting (with the piece count
displayed) if “AWP update” is set to “manual” or “automatic” in the Setup menu.
Manual updating can only be performed when the
Update soft key is displayed.
Reference sample updating must be completed before using an application program
from Application 3.
The
Update soft key is displayed when:
– the scale has reached stability (stability symbol displayed)
– the current piece count is less than double the original piece count
– the current piece count is less than 100
– the internally calculated piece count (e.g., 17.24) differs from the nearest whole number
(here: 17) by less than 0.3
Reference updating can be repeated several times with an approximately doubled
piece count.
§ To perform reference updating: Press the
Update soft key
§ Turn on the scale: Press e
> Sartorius logo is displayed, self-test is performed
§ Select the “Counting” program in the Setup menu: Press H
§ Select
Application parameters: press the q soft key 2+,
then the
O soft key once
§ Select
Application 1 (basic settings): press the O soft key
§ Select
Counting: Q or q soft key, repeatedly
§ Confirm
Counting: O soft key
Counting Average – average o Display accuracy
piece weight calculation +1 decimal place
+2 decimal places
Average piece Off
weight updating Manual
o Automatic
o = factory setting
see also the “Application Menu (Overview)” in the chapter entitled “Configuration”
§ Save settings and exit the Setup menu: Press the
oo soft key