Practical Example
Weighing in Components
Settings (changes in the factory settings required for this example):
Setup: Application parameters: Application 3: Formulation: Automatic storage: On, first value at stability
Setup: Application parameters: Application 3: Formulation: Minimum load for automatic storage: 100 digits
Setup: Application parameters: Application 3: Formulation: Evaluation mode, MR function: Final evaluation, print
Step Press key(s) Display/Output
(or follow instructions)
1. Turn on the scale and configure e
the settings as indicated above
2. Delete old formulation data c
3. Tare the scale t
4. Place the empty container Place load on the scale
on the scale (here: 180.59 g)
5. Tare the scale t
6. Weigh in the first component Place components
(here: 42.88 g) in container
7. Store components in the
M+ soft key
formulation memory
Scale is tared automatically
Component are printed out
automatically
isoTST
6
6
isoTST
isoTST
6
isoTST
6
--------------------
16.01.1997 14:04
Comp1 + 42.88 g