Purpose
With this function you can load weight
values and calculation results directly
from Application 1 (e.g., counting,
weighing in percent) or Application 2
(checkweighing, time-controlled func-
tions) into Application 3 (totalizing,
formulation, statistics).
Available Features
You can assign this function to the
fourth or fifth soft key (from the right),
i.e. F4 or F5
The soft key designation for this
function is:
M+
An Application 3 program (totalizing,
formulation or statistics) must be run-
ning so you can display and print the
result
Factory Settings
There are no optional parameters
Saving Values Manually (M+)
Preparation
G Turn on the balance: Press e
> Sartorius logo is displayed
G Select Extra function (F4) or Extra function (F5) in the Setup menu:
Press o
G Select
Application parameters: press the q soft key 2 +,
then the
O soft key
G Select
Extra function (F4) or Extra function (F5):
press the
q soft key 3 + (or 4 +), then the O soft key once
G Select
Man. store in app.3 memory (M+)
G Confirm Man. store in app.3 memory (M+)
See also “Application Parameters (Overview)” in the chapter entitled “Configuration”
G Save settings and exit the Setup menu: press the
oo soft key