Purpose
With this application you can determine
the density and volume of solid, pasty,
liquid or powdered samples.
You can use the “Density Determination”
application in combination with
a program chosen from Application 2
(e.g., checkweighing, timer functions)
and one from Application 3 (totalizing,
formulation, statistics) as well as the
extra functions.
63
Density Determination E
Available Features
Choose from 2 methods for determining
density of solids:
Buoyancy, or
displacement.
Density determination on pasty
or powdered samples using the pyc-
nometer method
Density determination on liquids using
the liquid density method
Choice of liquids for buoyancy:
Water
Ethanol
Other liquids (user-definable)
Reference values can be entered using
the numeric keys
Weight of sample in air (Wa)
Weight of sample in liquid, or weight
of reference liquid when using the
pycnometer (Wfl)
Weight of sample and reference
liquid when using the pycnometer (Wr)
Long-term storage of parameters:
Temperature
Buoyancy correction
Air density
Density of reference liquid
Expansion coefficient
Plummet volume
Wfl Wa Wr
Wfl
Wa
Wa
Wfl
Factory Settings
Method:
Buoyancy
Liquid causing buoyancy: Water
No. of decimals for display of vol.
density:
2 decimals
Printout: None
Soft Key Assignments
Wa Store weight of sample in
air
Wfl With liquid density,
buoyancy and displacement
methods:
Store weight of sample in
liquid
With pycnometer method:
Store weight of reference
liquid
Wr With pycnometer method:
Store weight of sample and
liquid
Start Start a new measurement
routine
Param. Toggle to parameter input
mode (depending on
method selected)
Densit Display the density
(the parameters set remain
effective for the next
measurement)
Weigh Display the weight
(the parameters set remain
effective for the next meas-
urement)
Vol. Display the volume
(the parameters set remain
effective for the next meas-
urement)

Questions and Answers: