93
Purpose
With this application program you can
have weight values and calculated val-
ues totalized and statistically evaluated.
The values determined for the evalua-
tion are:
average
standard deviation
variation coefficient
sum of all values
lowest value (minimum)
highest value (maximum)
difference between the minimum and
the maximum
You can use the “Statistics” application
in combination with a program chosen
from Application 1 (e.g., counting,
weighing in percent) and one from
Application 2 (checkweighing, time-
controlled functions) as well as with
the extra functions.
Available Features
Storage of weight values and calculated
values
Simultaneous storage of net and calcu-
lated values
Optional configuration in the Setup
menu for loading weight values and
calculated values either from Applica-
tion 1 (e.g., counting, weighing in
percent) or from Application 2
(checkweighing, time-controlled
functions)
Totalization memory for up to 65535
values
Simultaneous display in the text line
of transaction counter and, e.g., the
current total
Optional configuration in the Setup
menu for having the balance tare auto-
matically after a value is stored in the
totalization memory
Manual input of the number of
individual weighing operations (target
no. of operation
nDef). Result printed
and memory cleared after printout
of
nDef.
Optional configuration in the Setup
menu to add the current weight, with
display accuracy, to the current total by
pressing the
M+ soft key and generate
a printout of the result
Optional configuration in the Setup
menu for stability-dependent storage of
the measured value: Balance functions:
Stability range
Optional automatic storage of measured
values
Storage of measured value is indicated
by
Uu.
zz indicates that you can place a load
on the balance.
Minimum load threshold for automatic
storage
Press the
M- soft key to delete the last
value added to the totalization memory.
The transaction counter value is reduced
by one and a printout is generated.
Press the
MR: soft key for information
about number of transactions and the
current total. By configuring the Setup
menu, you can define whether the
information is displayed and printed,
or only printed, and whether the infor-
mation comprises an intermediate or
final evaluation (see the example)
In the Info window you can use the q,
l soft keys to choose which value
is displayed in the text line during
weighing
Printout of the end result depending on
the Application 1 or Application 2
parameters. Configure the Setup menu
to define which values are included on
the printout (Printout of individual
components)
Press
MR for a printout of an interme-
diate evaluation after each addition
or a final evaluation
A final evaluation is printed when the
statistics routine is ended by pressing
c, if no final evaluation was
generated by pressing
MR
Optional configuration in the Setup
menu to clear the totalization memory
and reset the transaction counter by
pressing c or after an evaluation
is printed out
Totalization data and transaction
counter data are stored in non-volatile
memory
Continue totalization after turning the
balance off and back on
Factory Settings
Automatic storage:
Off
Minimum load for automatic storage:
20 digits
Source of data for auto storage:
Application 1
Evaluated values: Net
Evaluation mode, MR key function:
Intermediate evalua-
tion, print
M+/M– function, then tare: Off
Printout of individual components:
Off
Balance function: Stability range:
2 digits
Printout: Application-defined output:
Print on request, then tare:
Off
Soft Key Functions
M+ Add weight values or applica-
tion values to the total in the
totalization memory.
The component or transaction
counter value increases by one
each time you press this key.
M- Delete the last value added
to memory. The transaction
counter value decreases by 1.
You cannot delete previous
values by
repeatedly pressing
this key.
MR Print or display an inter-medi-
ate or final evaluation
nDef Store the input number of
components
Statistics
N

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius LA Reference Series and is the answer not in the manual?

Sartorius LA Reference Series Specifications

General IconGeneral
BrandSartorius
ModelLA Reference Series
CategoryScales
LanguageEnglish

Related product manuals