Practical Example
Weighing in Components
Settings (changes in the factory settings required for this example):
Setup: App: Application 3: Formulation: Automatic storage: On, first value at stability
Setup: App: Application 3: Formulation: Minimum load for automatic storage: 100 digits
Setup: App: Application 3: Formulation: Evaluation mode, MR function: Final evaluation, print
Step Key (or instruction) Display/Output
1. Turn on the balance and configure e
the settings as indicated above
2. Delete old formulation data, if necessary c
3. Tare the balance w
4. Place the empty container Place load on the balance
on the balance (here: 180.59 g)
5. Tare the balance w
6. Weigh in the first component Place components
(here: 42.88 g) in container
7. Store components in the
M+ soft key
formulation memory
Balance is tared automatically
Components are printed out
automatically
8. Weigh in the next component Place components
(here: 50.80 g) in container
Components are stored in the
totalization memory at stability
and printed out
Balance is tared automatically
9. Repeat step 7 as required
10. Print final evaluation
MR soft key
(here: with total weight of all
components: 212.43 g)
11. Delete old formulation data, if necessary c
s
Comp2 + 50.80 g
s
--------------------
n2
Tot.cp+ 212.43 g
16.01.1997 14:10
--------------------