Setting the Balance Functions
(
BAL.FUNC.)
Purpose
This menu item enables you to
configure the balance functions, i.e.,
to meet individual requirements by
selecting predefined parameters in the
Setup menu. You can block access to
the menu by assigning a password.
Features
The balance functions are combined in
the following groups (1st menu level):
Calibration/adjustment
Adapt filter
Application filter
Stability range
Taring
Auto zero
Weight unit 1
Tare/zero at power on
Factory settings:
only wgh.
param. (only the balance functions)
Factory Settings
Parameters: The factory settings are
identified by the symbol “ο” in the list
starting on the next page.
15
Configuration
Preparation
Show available balance functions:
§ Select Setup menu:
press the o key
> SETUP is displayed
§ Select “Balance/scale functions":
press the O soft key
If you already assigned a password:
> The password prompt is displayed
$ If access is blocked by a password: enter
the password using the numeric/alpha-
betic keys.
$ If the last character of the password is a
letter: conclude input by pressing a
§ Confirm your password and have the
balance functions displayed: Press the
l soft key.
> Balance functions are displayed:
$ To select the next group: press the q
soft key (down arrow)
$ To select the previous item of a group:
press the Q soft key (up arrow)
$ To select the next sub-item within a
group: press the O soft key (right
arrow)
$ To select the previous group: press the
osoft key (left arrow)
$ To confirm: press the l soft key
Extra Functions
§ Exit the Setup menu: press the oo soft
key
> Restart your application
§ Print parameter settings:
When the balance functions are dis-
played, press r
> Printout (example)
Texts with more than 20 characters are
cut off
SETUP
BAL.FUNC.
---------------------
Calibration/adjustm
CAL/iso TST key fun
Internal cal./adju
Cal/adjustm seq
Cal. with adjustm au
.
isoCAL-function
Off
Start autom. adjus
isoCAL
Print GLP/GMP adju
Automatic if GLP is
selected
Parameter for exte
Wt. ID (W ID):
Cal./adjust.-wt:
5000.00 g
Adapt filter
Normal vibration
Application filter
Filling mode
Stability range
2 digits
Taring
After stability
Auto zero
Off
Weight unit 1
Grams /g
etc.