ISO/GLP-compliant Printout/Record
You can have the parameters pertaining
to weighing conditions printed before
(GLP header) and after (GLP footer) the
values from the weighing series.
These parameters include:
Date
Time at the beginning of a weighing
series
Balance manufacturer
Balance model
Model serial number
Software version
Lot number (weighing series no.)
Time at the conclusion of the weighing
series
Field for operator signature
Operating the Balance with an ISO/GLP-
capable Documentation Device (Printer)
ISO/GLP-compliant documentation
requires a computer with special soft-
ware. Contact Sartorius for a detailed
description for creating this software.
Setting:
Setup: Printout: ISO/GLP/GMP printout:
Always on
The record is output to a Sartorius
YDP03-0CE Data Printer or a computer.
End GLP printout:
G Press c
End GLP printout while application
is active:
This requires the following settings:
Setup: Device: Keys: CF function in
application: Clear only selected
applications
G Press c
> Text line: CF selected: clear application
G Press the
GLP soft key
--------------------
17.01.1997 16:12
SARTORIUS
Mod. LA4200S
Ser. no. 60419914
Ver. no. 01-35-18
ID 12345678901234
--------------------
L ID 12345678901234
nRef 10 pcs
wRef 1.35274 g
Qnt + 235 pcs
Qnt + 4721 pcs
S ID 12345678901234
Qnt + 567 pcs
--------------------
17.01.1997 16:13
Name:
--------------------
--------------------
17.01.1997 16:24
SARTORIUS
Mod. LA4200S
Ser. no. 60419914
Ver. no. 01-35-18
ID
--------------------
L ID 12345678901234
Internal calibration
Start: manual
Diff. + 0.006 g
Internal calibration
completed
Diff. + 0.000 g
--------------------
17.01.1997 16:25
Name:
--------------------
Dotted line
Date/time
Balance manufacturer
Balance model
Balance serial number
Software vers. (display and control unit)
Balance ID no.
Dotted line
Weighing series no.
Application initialization value
Application initialization value
Counting result
Counting result
ID for counting result
Counting result
Dotted line
Date/time
Field for operator signature
Blank line
Dotted line
Record of Internal
Calibration/Adjustment:
Dotted line
Date/time
Balance manufacturer
Balance model
Balance serial number
Software vers. (display and control unit)
Balance ID no.
Dotted line
Weighing series no.
Calibration adjustment mode
Beginning mode for calibration/adjustment
Difference after calibration/adjustment
Confirmation of completed
calibration/adjustment routine
Difference between current and target
values after calibration
Dotted line
Date/time
Field for operator signature
Blank line
Dotted line
116

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius LA Reference Series and is the answer not in the manual?

Sartorius LA Reference Series Specifications

General IconGeneral
BrandSartorius
ModelLA Reference Series
CategoryScales
LanguageEnglish

Related product manuals