Line Format (Data Output Format)
You can output the values displayed in the
measured value line and the weight unit
with or without a data ID code
Example: Without data ID code
+ 253 pcs
Example: With data ID code
Qnt + 253 pcs
Configure this parameter in the Setup
menu (Setup: Printout: Line format).
The output with data ID code has
16 characters; without data ID code,
22 characters.
118
Output Format With 16 Characters
Display segments that are not activated are output as spaces. Characters without a decimal
point are output without a decimal point.
The following characters can be output, depending on the characters displayed on the
balance:
Normal Operation
Position 12345678910111213141516
+ *DDDDDDDD *UUUCRLF
or – ........ ***
or * ********
*: Space
D: Digit or letter
U: Unit symbol
CR: Carriage return
LF: Line feed
Special Codes
Position 12345678910111213141516
******––******CRLF
or H H
or L L
or C
*: Space
– –: Weight
H: Overload
H H: Overload in checkweighing
L: Underload
L L: Underload in checkweighing
C: Calibration/adjustment
Error Codes
Position 12345678910111213141516
***Err**/###****CRLF
*: Space
# # #: Error code number