46
External Calibration/Adjustment* with
a
User-Defined Calibration Weight
First set either
Ext.
cal./adj.;user-defined
wt. or Selection mode (fac-
tory setting) in the Setup menu under
“Balance functions: Calibration/adjustment:
CAL key function.”
You can define a weight for calibration/
adjustment. External calibration/adjust-
ment must be performed with weights that
are traceable to a national standard and
that have error limits which are at least 1/3
of the required tolerance of the display
accuracy.
See page 44 for the external calibration/
adjustment sequence. For this example,
select external calibration/adjustment with
a user-defined weight.
The balance has a factory-set weight
value (see “Specifications”).
To reset a user-defined calibration
weight to the original factory setting:
$ Enter the factory-defined value
manually (see “Specifications”)
Define the Calibration Weight
Step Press key(s) Display/Output
(or follow instructions)
1. Select Setup menu o
2. Select “Balance/scale functions” O soft key
3. Select “Calibration/adjustment” O soft key
4. Select parameter for external q soft key 5 +
weight O soft key
5. Select “Cal./adj. wt.” q soft key
3000 = last setting selected
6. Enter calibration weight 5 0 0 0
(e.g., 5000 g)
and save
7. Save the calibration weight l soft key
8. Exit the Setup menu oo soft key